Residual solvents in pharmaceutical excipients, active pharmaceutical ingredients (APIs), and finished products are usually analyzed using gas chromatography (GC)-based techniques according to a pharmacopeial monograph, such as the United States Pharmacopeia’s (USP) chapter <467>. GC analyses are often slow, which limits sample throughput. Selected ion flow tube mass spectrometry (SIFT-MS) removes the rate-limiting chromatographic separation step, potentially offering faster sample analyses. This approach was demonstrated recently with the publication of an alternative SIFT-MS procedure which was successfully validated against the performance criteria in USP chapter <1467>. The present study expands upon the previous work by conducting a head-tohead comparison of GC-flame ionization detection (GC-FID) and SIFT-MS procedures. The results obtained in this cross-platform study demonstrated similar performance for the GC-FID and SIFT-MS procedures for linearity (R2 > 0.94 and 0.97, respectively) and repeatability (<17%RSD and <10%RSD). For accuracy and recovery, acceptance criteria (within 20%) were achieved for most compounds across the two drug products (SIFT-MS suffered fewer failures, possibly due to shorter wait times prior to analysis). Additionally, SIFT-MS analyzed samples over 11-fold faster than GC-FID, increasing daily sample throughput and reducing the time taken to determine the result. This study therefore suggests that residual solvent analysis using SIFT-MS may support workflow improvements for pharmaceutical manufacturers.